Waveforms and WFDB Export#
Isocenter ingests any DICOM instance that carries a Waveform Sequence
(5400,0100), alongside image data. Examples are 12-Lead ECG, General
ECG, Hemodynamic, and similar Waveform Storage IODs. Isocenter exports
these instances as PhysioNet WFDB records. The bridge is one-way: DICOM waveforms go out
as WFDB, for tools like Murmur Studio
to read. Isocenter does not read WFDB back in.
Quick start#
This section runs as written, as part of Isocenter’s test suite, over the
12-lead ECG bundled with pydicom. Start in a new, empty folder and copy
that file into input/:
import shutil
from pathlib import Path
import pydicom.data
Path("input").mkdir(exist_ok=True)
shutil.copy(pydicom.data.get_testdata_file("waveform_ecg.dcm"), "input")
from isocenter import Session
with Session("ecg_study.db") as session:
session.ingest("input")
# Write and load a configuration before auditing. An unconfigured
# session applies the floor policy (see below); the configuration
# is where you record the policy you actually want.
session.create_config("config.yaml")
session.load_config("config.yaml")
session.audit()
session.anonymize()
records = session.export("out", format="wfdb")
Prepare your own ECGs for a PhysioNet Challenge walks this path end to end, and then shapes each header the way the Challenge’s code reads it.
In a .py script, put the second block under
if __name__ == "__main__":
(why). Ingest starts worker
processes that re-import the script. A WFDB export saves the session
before it writes, as a DICOM export does. So the de-identified graph is
in the store when the block ends.
Each waveform instance becomes one WFDB record. The record is written
into the same directory tree that the DICOM exporter uses. So if you also
export format="dicom" into the same folder, a record’s .hea/.dat
files sit alongside that series’ .dcm files. export() returns the
path of each record’s header:
>>> records
['out/Subject_ANON_.../Study_..._..._.../Series_NoNumber_ECG_Series_.../ANON_..._0_1.hea']
>>> sorted(path.name for path in Path(records[0]).parent.iterdir())
['ANON_..._0_1.annotations.json', 'ANON_..._0_1.dat', 'ANON_..._0_1.hea']
The .hea file is the header, the .dat file holds the format-16
samples, and .annotations.json holds the cart’s findings, when the file
has any. The header’s signal lines end with each lead’s name:
>>> with open(records[0], encoding="utf-8") as header:
... [line.split()[-1] for line in header.read().splitlines()[1:13]]
['I', 'II', 'III', 'aVR', 'aVL', 'aVF', 'V1', 'V2', 'V3', 'V4', 'V5', 'V6']
The folder names are Subject_<Patient ID>, Study_<date>_<description>_<last 5 characters of the Study UID> and Series_<number>_<modality>_<description>_<last 5 characters of the Series UID>, each read from the exported values after
the configuration’s rules. A missing value becomes a placeholder such as
NoNumber or Series. A record name is the exported Patient ID, the Series
Number and the file’s Instance Number, joined by _. These parts are read
from the file’s own elements after the rules. A number that the rules
remove, or an empty one, reads 0, as does a number that the file never
had. (An empty Series Number reads NoNumber in the folder.) When two
records in one folder would share a name, the later ones get _2, _3,
and so on. The pseudonym, the shifted date and the UID suffixes differ in
your run, because all three are derived from a secret generated for each
store. That file carries two multiplex groups, so ingest also warns that
it kept group 0 and discarded 1 (see Limitations).
export(format="wfdb") returns the path of each record’s .hea file.
Each record that fails is logged, written to the audit log as an ERROR
row that names the instance, and counted in that export’s EXPORT row.
So the compliance report lists it and grades the run REVIEW_REQUIRED. A
partial export returns the records that did reach disk. When at least one
record was attempted and none was written, the call raises
io_handlers.ExportError after those rows, as a DICOM export does. Its
.failures names each record. A store can hold no waveform instances, or
only waveforms with no samples, and each waveform with no samples gets
its own DATA_LOSS row. An export of such a store attempts nothing and
returns [].
What is exported#
WFDB field |
DICOM source |
|---|---|
|
Sampling Frequency |
|
Derived from Channel Sensitivity |
|
Channel Sensitivity Units Sequence |
signal description |
Channel Source Sequence |
Signals are written as WFDB format 16 (16-bit, little-endian,
channel-interleaved) – the same layout DICOM already stores them in,
so no sample transcoding happens. The gain field is written in
spec-conformant header(5) form, gain(baseline)/units. If a
downstream tool mis-parses that field, check the tool’s parser first.
Lead names#
A lead whose Channel Source is coded in DICOM’s ECG lead context group,
CID 3001, is named by its conventional short name. That is the spelling
that KNOWN_LEAD_NAMES uses: I, II, III, aVR,
aVL, aVF, V1 to V9, V3R, V4R, V5R, X, Y, Z, and, in
MDC only, MCL1, MCL6, ES, AS and AI. Isocenter reads both
schemes that the context group has used. They are IEEE 11073 MDC
(2:1 is lead I), and the SCP-ECG codes that the group used before,
SCPECG (5.6.3-9-1 is lead I), which carts still write. The Coding
Scheme Designator is compared ignoring case, and the Code Value without
the whitespace around it ( 2:1 is 2:1).
Any other code, in any scheme, is written as its Code Value:
a derived or Frank lead
a calibration lead
a lead with no conventional short name (
-aVR,V2R,V6RtoV9R)a code under a local
99schemea source that is not a lead (a pressure or respiration waveform)
That Code Value is written as the file holds it. So a local code is
exported as the cart’s operator typed it. For the same reason, a rule of
your own that replaces Code Value (0008,0100) exports its dummy,
ANONYMIZED, as every signal’s name. Both are as they were before
1.0.2. Whether a code outside a published scheme should be written at
all is issue #973.
MDC 2:3 is Lead III or Lead V1. The 2008 version of the context
group (printed in PS3.16 2011) gave MDC 2:3 to both leads. The 2013
version, current in PS3.16 2026d, moved Lead III to 2:61. A cart built
between the two codes Lead III as 2:3, so the code alone cannot say
which lead it is. For this code only, Isocenter reads the Channel Source
item’s Code Meaning (0008,0104). When the Code Meaning says Lead III
(Lead III, lead iii or III), the signal is named III. Otherwise
the signal is named V1. The limit: a Lead III coded 2:3 whose Code
Meaning is empty, or says it another way (Ableitung III), is still
named V1. The same holds when a rule of your own replaces, empties or
removes Code Meaning (0008,0104). A 2008 Lead III is then named V1
again, and no row says so. (The basic profile has no such rule.) No
other code’s name depends on its Code Meaning, and nothing from the Code
Meaning is written.
Two signals of one record do not share a lead name. When two or more
signals of a record would carry the same name (compared ignoring case),
each one that got its name from the table above is
written as its Code Value instead. This applies in the .hea file and in
annotations.json alike. Two channels both coded MDC 2:2 are written
2:2 and 2:2, not II and II. The file says one code twice, and so
does the header. A name that did not come from the table (a Code Value
written as it stands, a kept Channel Label, a ch<N> token) is left as
it is.
The export writes one WARNING row for such a record. The row names the
channel numbers (from 1, as DICOM numbers them) and the lead name, and no
value from the file. So the compliance report grades the run
REVIEW_REQUIRED: a person should look at a record whose leads the
export could not name.
When present, Waveform Annotation Sequence (0040,B020) items –
cart-generated findings such as rhythm calls – are exported as
<record>.annotations.json, in the schema
Murmur Studio expects.
What is and isn’t de-identified#
Isocenter’s PHI scan is tag-gated, not content-based. A tag is
flagged only if your policy names it, by explicit phi_tags entries or
by a privacy profile expanded into them. A tag outside that set is never
inspected for what it contains, however identifying the text inside it is.
Private (odd-group) tags are removed whenever remove_private_tags is
True, the default, whatever phi_tags says. Setting it False does not
keep every private value. Large private binary values, which cart vendors
use routinely, are dropped at ingest (see
Private Tags).
Which policy runs depends on the configuration:
A bare
Session(), neverload_config()-ed, applies the floor policy: the basic profile plus three research defaults.The Quick Start above loads
privacy_profile: basic@2026c, the Basic Profile column of DICOM PS3.15 Annex E Table E.1-1 (2026c): 649 tags, 649 effective, every one reachable by the scan, including those inside sequences. The scaffoldcreate_config()writes carries the same three research defaults, so it applies the same 649 rules as a bare session.Your own configuration is layered on the floor, on a profile, or, with
privacy_profile: none, is the whole policy. Your explicit tags win over the base, andaction: KEEPopts a tag out.
Privacy Profile lists the floor’s research defaults and the profile’s departures from the table.
The basic profile empties Series Description (0008,103E) and
Study Description. That matters for waveform export because the series
description becomes a directory name: every .hea, .dat and
.annotations.json file lives inside it.
The exporter itself handles the three free-text surfaces specific to
waveform export, whatever the profile says. The PHI scan is tag-gated
(see above). So a profile entry alone protects only a session whose
policy carries it, and privacy_profile: none applies nothing. All three
are handled whichever configuration you run:
Channel Label
(003A,0203)– reaches the.heasignal-line description and theannotations.jsonleadfield only when it is a recognisable signal name, such as a standard ECG lead name. Anything else – including genuinely operator-typed text – is replaced with a positionalch<N>token instead of being written verbatim.Nis the zero-based channel index, not DICOM’s 1-based ChannelNumber –ch1is the second channel. The basic profile also removes Channel Label (PS3.15 Table E.1-1 gives itX). So on a configured or bare session, a channel with no coded Channel Source is namedch<N>even when its label was a known lead name. Give(003A,0203)action: KEEPto keep recognisable lead names.Unformatted Text Value
(0070,0006)– is omitted from thenotefield ofannotations.jsonby default. It routinely holds free-text clinical commentary, so its export is opt-in. To restore it, passsession.export(folder, format="wfdb", include_annotation_text=True).(0070,0006)is also in the basic profile. So a configured session that passesinclude_annotation_text=Trueexports the profile’s remediated value. That value is the text dummyANONYMIZEDin the exported DICOM, as Table E.1-1’sDasks. The bridge reads the dummy as no text, so the finding carries nonote, exactly as when the value was emptied.Concept Name
(0040,A043)– the annotation’s Code Meaning reachesannotations.jsonaslabel, and its scheme-qualified Code Value ascategory. Both reach it only when the Coding Scheme Designator(0008,0102)names a published vocabulary. Examples are SNOMED CT, LOINC, DICOM’s ownDCM, IEEE 11073MDCand SCP-ECG (SCPECG, which ECG carts use for measurements such as QRS Onset). A coded finding is unaffected.SCT:164889003still arrives with its label"Atrial fibrillation", because SNOMED defined that term, not an operator.For a site-defined scheme the cart populates Code Meaning with typed text instead, so
labelis omitted andcategorycollapses touncoded. The finding itself survives – kind, sample positions and lead are untouched. This is because a reviewer seeing fewer marks than the record carried, with nothing saying any were withheld, is a worse failure than seeing them unnamed. What is lost is the name and the ability to group two site-defined annotation types apart from each other.DICOM reserves designators beginning
99for locally defined schemes, so a99...value is never recognised, however conformant it looks.include_annotation_text=Truerestores both fields, exactly as it does fornote. That flag is the protocol’s voice here. A study whose auditor determined that site-defined annotation labels may be released says so by passing the flag.
All three are safe by default. No PHI tag configuration is required to
get this behaviour, and it applies even to a bare Session().
annotations.json’s source field is producer provenance only. It holds
the running Isocenter version plus Manufacturer (0008,0070), for example
isocenter/1.0.0 (Mortara Instrument, Inc.). It does not read Device Serial Number
(0018,1000) or any other equipment identifier.
Record timing in the .hea file combines two independently
sourced parts. The date comes from study.study_date, which is
shifted by anonymize(). That is the same per-patient date shift that is
applied to every date tag under a SHIFT or JITTER rule. The
time-of-day comes from the instance’s own timestamp tags: Acquisition
DateTime (0008,002A) when present, else Study Time (0008,0030). The
date shift never changes it. This is deliberate: time-of-day alone is not a
Safe Harbor identifier. If you export without running anonymize() at
all, the date is real too – exactly like every other un-remediated
field in Isocenter.
On the documented Quick Start path, both instance-level timestamp tags
above are in the Basic profile ((0008,002A) and (0008,0030)). So a
configured, anonymized session has no real time-of-day left to write.
Isocenter does not substitute a fake 00:00:00 in that case. When
study.study_date is real but no real time-of-day is available, the
record line’s start time/date fields are omitted entirely. This is a
deliberate limitation, not an oversight. header(5) does not support
a date-only start time. PhysioNet’s own spec, and wfdb-python’s
reference reader/writer, both treat base_date as depending on
base_time being present. So the record line’s choice is between
omitting both fields and fabricating a time. Isocenter omits both.
The date is not lost, though, because it is useful for ordering records
within a cohort. It is written instead as a single comment line, in the
same DD/MM/YYYY format that the record line’s own date field would have
used:
# de-identified start date: DD/MM/YYYYwhenanonymize()shifted the study date# start date: DD/MM/YYYYwhen the study date was not shifted (an export withoutanonymize(), or a date the policy keeps)
That second form is the real date. The comment follows the signal lines,
where wfdb’s own writer puts comments. So a reader that takes lines
1..n_sig as the signal lines reads the signals and nothing else. (The
Moody Challenge’s helper code is such a reader.) A consumer that reads
the record line alone sees no timing at all. A consumer that also reads
comments gets the date, labelled by whether it was shifted.
Beyond content, the export path itself avoids two structural PHI paths a WFDB writer could otherwise open:
No
#comment lines are written, except the start date. WFDB readers render header comments verbatim, and MIT-BIH convention places age, sex, and diagnosis there. Isocenter never writes one for content. The one exception is the start-date line described above (# de-identified start date:or# start date:). That line is a computedDD/MM/YYYYstring, written through the same sanitizer as every other field, and never operator-typed text.Record names are built from the exported Patient ID, series number and instance number. That Patient ID is the pseudonym once
anonymize()has run, and the source Patient ID before it, or under aKEEPrule on Patient ID. A record name never holds the patient’s name or other free text. Lead identity in the header prefers the coded channel source over the free-text label wherever a coded source exists.That preference is a likelihood argument, not a filter. A conformant coded source is far less likely to carry operator-typed text than a free-text label. But a coded value other than a CID 3001 lead code is written as it stands, not run through the lead-name allowlist or any content check. A non-conformant source can still put arbitrary text there. Both the
.heawriter and the Murmurannotations.jsonbridge remove line-break characters from that text regardless. So the text can’t forge a.heacomment line, but the text itself is trusted, not filtered.
DICOM to DICOM round trips#
Exporting with format="dicom" writes the waveform samples back. The
bytes are copied from the sidecar verbatim, not re-encoded from the
decoded array. So a round trip is byte-exact, and the exported
Waveform Sample Interpretation (5400,1006) cannot disagree with the
samples it labels.
That verbatim copy is deliberate. Isocenter decodes US/SB/UB to
int16 internally, and rebases US by 32768. A re-encode from that array,
while (5400,1006) still says US, would shift every value by 32768
without anything raising. Nothing in the pipeline mutates waveform
samples – unlike pixels, which redaction burns into. So there is
nothing a re-encode could add, and the mismatch is made structurally
impossible instead of merely tested against. De-identification is
unaffected. Waveform PHI lives in tags (channel labels, annotation
concepts), which are remediated in the object graph and rebuilt into the
exported dataset as usual.
Because no decode happens on this path, companded audio (MB/AB)
round-trips through DICOM export even though WFDB export refuses it.
Byte order. Ingest converts a big-endian source’s samples to
little-endian, by the Waveform Bits Allocated it declares, and both exports
write them under a little-endian transfer syntax. So the round trip is
exact against the samples as stored, not against the source file’s bytes. A
value whose byte order ingest could not settle whole is kept as read, with a
WARNING row naming the tag.
If an instance carries a Waveform Sequence but no samples reached the sidecar, the export logs a warning rather than writing a structurally-plausible empty record in silence.
Limitations#
Format 16 only. The exporter also does not support: WFDB ingest,
.atr annotation output, or mu-law/A-law companded audio sample
interpretations.
Multi-rate records are truncated at ingest, not at export. Each
Waveform Sequence (5400,0100) item is a multiplex group with its own
sampling frequency and channel set. That is how DICOM carries ECG at
500 Hz alongside respiration at 25 Hz. Isocenter reads group 0 and
discards the rest, so groups 1..n never enter the object graph. Neither
their samples nor their sequence items are in the session, in the
sidecar, or reachable by any export format or API. What is written is a
conformant single-group record. The discard is announced: a warning
naming the number of groups dropped, plus a DATA_LOSS row in the audit
log. Multi-group support is not yet available
(#277). A store written
before 0.9.1 can still hold the discarded groups’ items.
That row is scoped SIGNAL, and it does move Validation Status. A
session that discarded a multiplex group grades REVIEW_REQUIRED, not
PASS. The tag is standard – Waveform Sequence (5400,0100) is an
even group. But what was discarded is acquired signal that was in the
source and is not in the export. That loss is not routine the way a
dropped overlay is. Routine standard-group losses keep the STANDARD
scope and keep grading PASS.
Annotations naming a discarded group are dropped, not re-pointed.
Referenced Waveform Channels (0040,A0B0) identifies a mark by a
(multiplex group, channel) pair, and DICOM numbers both from 1. PS3.3
C.10.10.1.1 “Referenced Channels” defines the first value of each pair as
the ordinal of the Waveform Sequence (5400,0100) Item. Its worked
example writes an annotation that covers the entire first multiplex group
plus channels 2 and 3 of the third as 0001 0000 0003 0002 0003 0003.
Ordinal 1 is therefore the group Isocenter keeps. (The 0000 in that
example is the same section’s rule that a channel number of 0 means every
channel in the group. Isocenter honours that rule by emitting the mark
with no lead.) A mark that names any other group has no place in the
exported record. Its sample positions index that group’s samples, and
its lead is that group’s channel. Such a mark is left out of
annotations.json, not resolved against the surviving group. That
resolution would give a plausible lead name at a plausible sample
position, both belonging to a signal that is not in the record.
A mark that names a channel of the kept group which the record has no
signal for is kept, with no lead. That case is a file with more Channel
Definitions than NumberOfWaveformChannels. The .hea has one signal
line per sample column. So a mark on a channel past the last column names
a lead that no signal of the record carries. Its sample position is still
on the record’s axis, so the mark stays.
The same filtering happens on the object graph at ingest. So a DICOM
export does not carry a Waveform Annotation Sequence (0040,B020) item
whose reference names a group that is not in the file. The filter works
on (group, channel) pairs before it drops items. An annotation that
names all of group 1 plus a channel of group 3 keeps its surviving pairs,
and an item goes only when every pair it named is gone. Surviving
ordinals are never renumbered. The ordinal is positional, so a
renumber after a discard would make the file internally consistent and
wrong relative to the source, with no way to tell afterwards. An
annotation with no (0040,A0B0) at all is untouched. The attribute is
Type 1C, and its absence means the mark applies to the whole waveform.
That drop is announced the same way the group discard is: a warning, and
one DATA_LOSS row per instance. The row names how many annotations were
dropped and which groups they referenced. It is one row, not one per
mark, so a cart that marks forty beats on a discarded group does not fill
the report’s Data Loss section with forty near-identical lines.
That row is scoped STANDARD, although the group discard itself is
scoped SIGNAL. An annotation is a mark about the signal. The loss of
the samples already moves Validation Status through the group’s own row,
and to grade the marks too would count one loss twice. A record with two
DATA_LOSS rows – one for the groups discarded, one for the marks that
referenced them – is the expected shape. The two rows report different
losses. Both rows are written at ingest. The WFDB export repeats the
drop, with its own row, only for a graph that never passed through
ingest.